FLOW CYTOMETRY · PANEL DESIGN
Spectra viewer and panel planner
Compares fluorophore excitation and emission spectra against the lasers and detectors of a specified cytometer, and estimates spillover and spectral similarity for a panel.
Edit lasers and detectors
One row per laser. Bandpass filters are entered as centre/width in nanometres, separated by commas, for example 450/50, 525/50. The presets are typical configurations and should be adjusted to the filters fitted to the instrument in use.
Spectra
Excitation Emission Detector bandpass
Panel
| Fluorophore | Marker | Laser | Peak detector | Emission peak | Brightness | Remove |
|---|
Select fluorophores from the list to build a panel.
Overlap
Methods
Signal. For each fluorophore and detector, the predicted signal is the fluorophore's relative excitation at the detector's laser line multiplied by the fraction of its emission within the detector's bandpass. Filters are modelled as ideal rectangular bandpasses.
Spillover. The signal of each fluorophore in the peak detector of every other panel member, expressed as a percentage of the signal in its own peak detector. The table follows the layout of a compensation matrix, read across each row.
Similarity. The cosine similarity of two fluorophores' signatures across all detectors on the instrument, from 0 (no overlap) to 1 (identical). Pairs above approximately 0.98 cannot be resolved. The condition number of the similarity matrix indicates how well the panel as a whole can be unmixed; lower values are better.
Limitations. Results are calculated from reference spectra, not from measurements on a specific instrument. Laser power, detector sensitivity, antigen density, tandem dye lot and fixation all affect measured signal. Brightness tiers are indicative. Panels should be confirmed experimentally with single-stained controls.